Abstract

A reliable and well-characterized dosimetry system which is traceable to the international measurement system, is the key element to quality assurance in radiation processing with cobalt-60 gamma rays, X-rays, and electron beam. This is specifically the case for health-regulated processes, such as the radiation sterilization of single use medical devices and food irradiation for preservation and disinfestation. Polyethylene is considered to possess a lot of interesting dosimetric characteristics. In this work, a detailed study has been performed to determine the dosimetric characteristics of a commercialized high-density polyethylene (HDPE) film using Fourier transformed infrared spectrometry (FTIR). Correlations have been established between the absorbed dose and radiation induced infrared absorption in polyethylene having a maximum at 965 cm−1 (transvinylene band) and 1716 cm −1 (ketone-carbonyl band). We have found that polyethylene dose-response is linear with dose for both bands up to1000 kGy. For transvinylene band, the dose-response is more sensitive if irradiations are made in helium. While, for ketone-carbonyl band, the dose-response is more sensitive when irradiations are carried out in air. The dose-rate effect has been found to be negligible when polyethylene samples are irradiated with electron beam high dose rates. The irradiated polyethylene is relatively stable for several weeks after irradiation.