Abstract

BackgroundThe determination of the osmolality of aqueous samples using a freezing point osmometer is a well-established, routine laboratory method. In addition to their use in clinical and pharmaceutical laboratories, freezing point osmometers are also employed in food testing laboratories. One application is the determination of the osmolality of milk. Although cow's milk is a natural product whose water content is approximately 87%, the osmolality of milk is a significant value when the milk is collected from a larger population of animals. This value is used in milk processing to control the water content, based on the German Food Control Regulations for Milk.ResultsMeasurement of the freezing point and osmolality of milk samples was performed with a Knauer Semi-Micro Freezing Point Osmometer. Osmolality was measured for the untreated milk samples and following their dilution (by volume) with 10% and 50% water. The measurements were made after 1, 4 and 7 days to evaluate changes over time. All measurement values for the undiluted milk were spread over a small interval with an average of 271 mOsmol/kg. After mixing the milk samples with 10% water, the average decreased to 242 mOsmol/kg, while mixing with 50% water resulted in an average osmolality of 129 mOsmol/kg. There was no significant change for the osmolality within the 7 days (measurements from days 1, 4 and 7).ConclusionThe results observed demonstrate clearly that the additional water content of milk can be determined easily using a freezing point osmometer. Milk samples that contain additional water have a significantly decreased osmolality, corresponding to an increased freezing point. The effect on osmolality of ageing the milk samples could not be determined in this study's time-dependent measurements.

Highlights

  • The determination of the osmolality of aqueous samples using a freezing point osmometer is a well-established, routine laboratory method

  • The freezing point osmometer is in use in the food laboratory, where its main applications are in the determination of alcohol content in beverages and quality control of other additives [3]

  • Chemistry Central Journal 2008, 2:6 http://journal.chemistrycentral.com/content/2/1/6 case), in all these applications the solute is that which is determined by osmometry

Read more

Summary

Introduction

The determination of the osmolality of aqueous samples using a freezing point osmometer is a well-established, routine laboratory method In addition to their use in clinical and pharmaceutical laboratories, freezing point osmometers are employed in food testing laboratories. Cow's milk is a natural product whose water content is approximately 87%, the osmolality of milk is a significant value when the milk is collected from a larger population of animals. The determination of aqueous samples' osmolality using a freezing point osmometer is a well-established, routine laboratory method. The freezing point osmometer is in use in the food laboratory, where its main applications are in the determination of alcohol content in beverages and quality control of other additives [3]. Its additional water content is the most important factor for the determination of milk osmolality [4,5]

Objectives
Results
Conclusion