Abstract

This study aimed to evaluate the fracture resistance of space maintainers (SMs) produced using 3D-printable materials (metal, resin and polyetheretherketone [PEEK]) after thermal aging and compare them with conventional space maintainers. A standardised typodont model for paediatric dentistry was utilised, and band and loop space maintainers were designed digitally using computer-aided design (CAD) technology. Four groups were established: Conventional, 3D printed metal, 3D printed resin, and 3D printed PEEK. Fracture resistance was assessed after 10,000 thermal cycles, simulating oral conditions. Fracture tests were conducted using a universal testing machine, applying vertical force to the band and loop junction until fracture. Statistical analyses were performed using one-way ANOVA and the Tukey HSD test (P<0.05). Although the fracture resistance values showed that metal, resin, and PEEK 3D printed band and loop space maintainers can be acceptable clinically, the permanent resin may be preferable to printable material because of their aesthetic properties.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.