Abstract
This prospective clinical study evaluated the incidence of instrument fracture observed after single-file root canal treatment of molars using WaveOne Gold instruments. Three standardized, experienced and calibrated specialists treated 750 maxillary and mandibular molars with curvatures less than 45° (2691 root canals) over a 12-month period. All the treatments were performed in a single session. A total of 1104 WaveOne Gold instruments were used, including 38 small, 750 primary, 228 medium and 88 large instruments. Intracanal procedures were performed according to the manufacturer's recommendations, and each instrument was used in a single clinical case. The instruments were examined after their removal from the canal, under an operating microscope at 8× magnification. No fractures were observed in any of the 1104 instruments used. No fractures of WaveOne Gold reciprocating instruments occurred during root canal preparations performed in maxillary and mandibular molars with curvatures less than 45° when used strictly according to the manufacturer's recommendations and applied in a single clinical case.