Abstract Many commercial linear polyethylenes have very broad distributions of molecular weight. The high molecular weight fractions often extend beyond the highest molecular weight calibration standard of GPC. For this reason the reliability of information obtainable from GPC has been examined with attention to the average molecular weights. Calibration range is a serious limitation for the accurate determination of the - weight-average and the higher averages of molecular weight. Uncertainty in the baseline at the high molecular weight region, however, does not produce a significant error. With a four-column GPC having 103 to 107 A nominal capacity, improved resolution is needed in the high molecular weight range. In order to examine the resolution and to improve the calibration, a polyethylene standard of ca. 3–4 million molecular weight is required. With the present limitation of GPC the greatest amount of information can be obtained by examining and intercomparing the cumulative distribution curves. ...

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.