Abstract
Transvenous lead extractions (TLEs) for cardiac implantable electronic device complications often encounter difficulties with strong adhesions to the myocardium or vessels. In this report, we introduce a novel "Four-Stage Rocket" technique for effective TLE in cases where conventional methods fail. Two challenging cases where conventional TLE methods failed were treated using a combination of four devices: Needle's Eye Snare, Agilis NxT Steerable Introducer, GlideLight Laser sheath, and GORE® DrySeal Flex Introducer sheath, employed via the inferior vena cava. The "Four-Stage Rocket" technique successfully detached firmly adhered leads near the tricuspid valve annulus, where the traditional superior vena cava approach was inadequate. The "Four-Stage Rocket" technique offers a potential alternative in complex TLE cases, aligning the laser direction with the adhesion detachment and reducing the tissue damage risk.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.