Abstract
EPR spectra of alanine dosimeters irradiated by gamma-rays and by alpha-particles were measured at different microwave power. The spectra were decomposed into three components corresponding to different stable radiation-induced radicals. Different ratio of the components was obtained for gamma- and alpha-irradiated alanine. The observed effect can be used for discrimination of the contribution of alpha- and gamma-radiation in mixed radiation field. That may be useful for control of dose at phantom experiments with neutron boron capture therapy, performed in mixed neutron-gamma field in which alpha-particles are emitted through neutrons capture reaction with boron.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.