Abstract

Simple and precise stability indicating RP-HPLC method for the quantitative determination of tizanidinehydrochloride (TZN) in pharmaceutical dosage form was developed and validated. The separation was achieved using Atlantis C18 column (250 x 4.6 mm, 5µm) at 25 °C using a mobile phase containing 20mM KH2 PO4 (pH 3.5):methanol in the ratio of 30:70 V/V at 0.8 mL/min flow rate and 315nm detection wavelength. The retention time for TZN was 3.7 min and showed linearity in the 5-40µg/mL range with R2 >0.99. The drug was subjected to acid/ base hydrolysis, oxidative and thermal degradation to establish stability indicating method. The method was validated as per ICHQ2 (R1) guidelines. The method was accurate, precise, and specific for TZN estimation. In stress studies, the drug was found to be stable to acid/alkaline hydrolysis, oxidation, and thermal degradation conditions. Thus, the reported method can be used as a stability-indicating method for quality control and routine analysis of TZN.