Abstract

A convenient fluorimetric method for the routine determination of aluminium in serum has been developed using lumogallion [4-chloro-3-(2,4-dihydroxyphenylazo)-2-hydroxybenzene-1-sulphonic acid]. Losses of aluminium during deproteinisation of the serum were prevented by treatment with a combination of 20 or 30% m/V trichloroacetic acid (TCA) and 5% m/VTCA. Iron(III) was removed by extraction into chloroform with capriquat (methyltrioctylammonium chloride) as an Fe3+-lumogallion-capriquat ternary complex. The interference from Cu2+ was eliminated by using thiosulphate as a masking agent. The detection limit was 3.6 ng ml-1 and the calibration graph was linear up to 1.4 micrograms ml-1 of aluminium. Using the proposed method, the average concentration of aluminium in the serum of healthy subjects was found to be 6.8 ng ml-1, in agreement with values reported in the literature.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.