Abstract

AbstractAn immersed flowcell has been developed for fluorescence detection in capillary electrophoresis (CE) by modifying an ordinary HPLC spectrofluorimetric detector flowcell. The minimum detectability obtained for riboflavine was about 4 fmol. The use of the flowcell to measure a fluorescence spectrum by the stopped‐flow method of shutting down the DC high voltage during a wavelength scan, has also been achieved successfully.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.