Abstract

The role of the upper airway (the breathing passage above the trachea) in maintaining the normal junction of the respiratory system has been suggested by previous investigators. During a tracheotomy the upper airway is by-passed by a prosthetic metal or plastic tube which is placed into the trachea through the neck. In order to determine which, among the most commonly used tracheotomy tubes, most closely simulate the flow resistance of the adult human upper airway, a series of varying flow rates were passed through different sized tubes. Pressure drops were recorded and resistance values were thereby determined. Our data was compared with previously determined values for flow resistance of the adult human upper airway. Resistance related to turbulent and laminar flow was considered. On the basis of our data we have suggested that large caliber tracheotomy tubes be used in adult patients in whom the prolonged need for a tracheotomy is anticipated.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.