Abstract
BackgroundPerforming reproducible flow-mediated dilatation (FMD) measurements can be challenging, especially among inexperienced operators. Our aim was to present the detailed learning curve of a novice operator.MethodsFollowing a one-week basic training period, the operator performed duplicate measurements on 6–8 individuals per week, for six weeks. The operator followed the recommendations of the most recent guideline. Duplex ultrasound was used for the simultaneous and continuous measurement of brachial artery diameter and local blood flow velocity. Following a 1-min recording of baseline diameter (D), FMD was measured after a 5-min period of occlusion of the proximal forearm vessels. Inter-session coefficient of variation (CV) values for D and FMD were calculated for each week.ResultsThe number of volunteers assessed each week were the following: 8, 7, 7, 6, 7, 7 individuals. CV values for both D and FMD exhibited a decreasing trend over the training period, ending at 1.73% for D and 14.24% for FMD at week 6. These CV values are within the range outlined in the most recent FMD guideline for proficiency.ConclusionWithin a reasonable timeframe, and with careful adherence to measurement guidelines, the attainment of sound reproducibility in FMD measurements by a novice operator is feasible.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have