Abstract
ABSTRACTThe development of a simple and sensitive on-line flow injection preconcentration system is reported using a polyamino-polycarboxylic acid chelating resin for the automated determination of cadmium(II), chromium(III), copper(II), and lead(II) in natural waters and urine by flame atomic absorption spectrometry (FAAS). The adsorbent showed a high sorption affinity for the metal ions at pH values from 5.0 to 6.0. The retained analytes were efficiently eluted by 1.5 mol/L HNO3. The effect of chemical and hydrodynamic parameters such as pH, selection of eluent (type, volume and concentration), and sample/eluent flow rates were studied, providing optimum conditions for the analysis of real samples. For a preconcentration time of 120 sec, the sampling frequency was 22/h and the enhancement factors were 105, 49, 106, and 98 for cadmium(II), chromium(III), copper(II), and lead(II), respectively. The detection limits were between 0.1 and 1.6 mg/L, while the precision was less than 3.3% as the relative standard deviation for all metal ions. The accuracy of the method was evaluated by the analysis of certified reference materials and fortified environmental and biological samples.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.