Abstract
A simple and sensitive flow injection on line separation and preconcentration system coupled to hydride generation atomic fluorescence spectrometry (HG‐AFS) was developed for ultra‐trace bismuth determination in water and urine samples. The preconcentration of bismuth on a nylon fiber‐packed microcolumn was carried out based on the retention of bismuth complex with Bismuthiol I. A 15% (v/v) HCl was introduced to elute the retained analyte complex and merge with KBH4 solution for HG‐AFS detection. Under the optimal experimental conditions, an enhancement factor of 20 was obtained at a sample frequency of 24/h with a sample consumption of 13.0 ml. The limit of detection was 2.8 ng/l and the precision (RSD) for 11 replicate measurements of 0.1 µg/l Bi was 4.4%.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.