Abstract

Purpose The flow injection distillation (zinc acetate solution) was changed to be 5% phosphoric acid in the method of determination of cyanide. Application Cyanide in groundwater was measured by FIA6000 + automatic flow injection analyzer directly. Results In the 2.0-100.0μg/L range, the linear correlation coefficient was greater than 0.9, the recovery was 86.1% -100.8%, relative standard deviation 1.3% -3.8%. Conclusion This method not only reduces the white crystals [Zn (OH)2] on the pump tube blockage, making the operation more simple and efficient, but also for the detection of a large number of groundwater samples.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.