Abstract

Copper was determined by flow injection spectrophotometry at 370 nm after extraction of copper(II) bis(acetylacetone)ethylenediiminate chelate into chloroform. The carrier stream was deionized water and the reagent streams were acetate buffer (pH 5.0) and 0.2% (w/v) bis(acetylacetone)ethylenediimine solutions. The sampling rate was 20 h −1. The calibration graph was linear up to 100 μg ml −1 copper, based on injection volumes of 250 μl. The relative standard deviation ( n = 10) was 1.95% for 20 μg ml −1 copper. The system has been applied to the determination of copper in copper-based alloys and pig feeds.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.