Abstract
Based on the flow injection principle and employing a nitrate electrode situated in a flow-through cell, a simple, reliable and sensitive method for the automated determination of nitrate and its use for analysing soil extracts, waste waters, fertiliser solutions and air samples are described. At a sampling rate of 90 samples per hour a standard deviation of only 1–2% is typical. A new approach to identifying and correcting for the presence of interfering species in potentiometric measurements is discussed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.