Abstract

The flow injection principle is used with novel design of a flow cell, in which the ion-selective and reference electrode are incorporated. The reproducible mixing of the sample with the carrier stream of electrolyte and the precise timing of potentiometric measurements in the continuously moving stream are the most important parameters influencing the quality of direct potentiometric measurements. Methods for the determination of nitrate and potassium in soil extracts, and sodium and potassium in blood sera are described. At sampling rates of 125 determinations per hour, a standard deviation of only 0.8 % is typical. Simultaneous measurements of two ions on one stream are described; advantages and drawbacks of the use of ion-selective electrodes in continuous flow measurements are discussed; and the new possibilities which the flow injection method offers are contemplated.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.