Abstract

This study aims at an electrochemical paper-based device for the selective and sensitive detection of exosomes as an economic and flexible platform. Using a laser cutter, a three-electrode platform was patterned on a hydrophobic paper, and then synthetic carbon and silver ink were screen-printed on the paper. Anti-human CD9 antibody was covalently attached to the surface of a carbon ink electrode enriched with graphene oxide aerogel/CoFe2O4@chitosan nanocomposite. The immobilization, incubation and binding time of the capture antibody were investigated as effective parameters. Under optimum conditions, the differential pulse voltammetry (DPV) was used to establish a calibration plot that covered 500 to 5 × 106 exosoms μl−1 with a detection limit of 100 exosome μl−1. This method was implemented successfully for the detection of exosomes in a serum sample. As the results show, the fabricated biosensor is capable of detecting exosomes for the early diagnosis of cancers.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call