Abstract

This study reviewed 863 alerts generated from the infusion of anticoagulants in 355 patients from October 2003 to January 2005. Alerts were generated by smart infusion technology pumps and recorded in the devices’ memory. The most common alerts were underdose alerts (59.8%), followed by overdose alerts (31.3%) and duplicate drug therapy alerts (8.9%). In response to the alerts, users’ most frequent action was to cancel (46.5%) or reprogram (43.1%) the infusions. The highest percentage of alerts occurred from 2 to 4 p.m. During the study, there were 4 infusion rate errors, compared with 15 in the immediately preceding 16-month period. In conclusion, smart infusion technology intercepted keypad entry errors, thereby reducing the likelihood of intravenous anticoagulant overdose or underdose. Dose or infusion rate programming during intravenous anticoagulation is an important targets for medication safety interventions.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.