Abstract

A flameless atomic-absorption method for determination of phosphorus is developed, employing zirconium treated graphite tubes. The great improvement of the sensitivity with these tubes eliminates the necessity of adding La(NO3)3. The sensitivity does not vary from tube to tube and its maximal value is attained within the first 2–3 firings. The atomization signal is practically the same within the range of 2,600–2,800° C when the rise of the temperature is high (> 1,500° C/s). The best signal to noise ratio is obtained at 2,600° C. The detection limit is 0.05 ppm with injection of 50 μl. The method is applied to the analysis of a standard steel sample.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.