Abstract
ABSTRACTThe general purpose powder diffractometer (GPPD) is the first powder diffractometer at the China Spallation Neutron Source (CSNS). Its aim is to provide a high-resolution, good-intensity instrument for users to study crystallographic and magnetic structures. The best resolution will be better than 0.2% at high-angle detector bank. The neutron flux at the sample position will be at the order of 107 n/s/cm2. The joint beam commissioning among accelerators, target station and instruments started in the CSNS on November 1, 2017. The neutron beam shutter for the general purpose powder diffractometer (GPPD) was opened for the first time. Initial measurements and standard sample diffraction experiments are reported, which are used to characterize the instrument performance. It is shown that the results agree well with the design goal. During the commissioning, a series of user experiments have been conducted. Two user experimental results have already been published.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.