Abstract

Abstract Introduction: Mandibular fractures are common injuries during maxillofacial trauma, and currently, open reduction and internal fixation are considered gold-standard treatments. There is a wide discussion about which plates give the best outcomes. Hence, we are conducting a biomechanical comparison of two plates for mandibular symphysis and angle fracture with finite element analysis (FEA). The aim of this study was to do a comparative study of FEA between the conventional and our new modified three-dimensional (3D) strut miniplate in mandibular fractures at symphysis and angle regions. Materials and Methods: Finite element models of symphyseal and angle fractures of the mandible were developed. Each fracture model was then realigned and fixed by the conventional method 2.0 mm system, and our modified 3D strut plating method 2.0 mm system followed by the analysis of various stresses developed in plates and mandibular fracture area after application of load was observed in the study. Results: The modified 3D strut plating system with 2.0 mm miniplates is significantly better in preventing displacement of fracture segments by better distribution of forces compared to the conventional plating system. Rest of the parameters were within the permitted limits. Discussion: Modified 3D strut plating method was reasonably effective and superior in managing force-displacement compared to the conventional method of fixation for comminuted and unfavourable mandibular symphyseal fracture and angle fracture.