Abstract
This final report describes the results obtained in a key comparison of primary vibration calibration facilities. Two piezoelectric standard accelerometers (one single-ended as well one back-to-back type) were calibrated first at Central Office of Measures (GUM), acting as pilot and linking laboratory, and Bulgarian Institute for Metrology (BIM). The results of the BIM obtained in this comparison were linked to the CCAUV.V-K2 key comparison through the joint participation of the GUM. The degrees of equivalence were computed for BIM with respect to the CCAUV.V-K2 key comparison reference value.Main textTo reach the main text of this paper, click on Final Report. Note that this text is that which appears in Appendix B of the BIPM key comparison database kcdb.bipm.org/.The final report has been peer-reviewed and approved for publication by the CCAUV, according to the provisions of the CIPM Mutual Recognition Arrangement (CIPM MRA).
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.