Abstract

We recently described a lightweight, low-power, waterproof filter fluorometer using a 180° backscatter geometry for chlorophyll-a (chl-a) detection. Before it was constructed it was modeled to ensure it would have satisfactory performance. This manuscript repeats the modeling process that allows the calibration slope and detection limit for a fluorescent analyte in water to be estimated from system component performance and conventional spectrofluorometry alone. These values are validated by comparison to the experimental result of calibration from the completed instrument. Our model yields a calibration slope of 8.22 mV-L/µg for dissolved chl-a, consistent with the experimentally measured slope of 8.21 mV-L/µg. The detection limit modeled from this slope and an estimate of the baseline noise of the instrument was 0.15 µg/L chl-a, while the measured detection limit using real blank samples was 0.18 µg/L, in 0.1 s differential measurements.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.