Abstract

The design of a nonselective radiometer that functions on the basis of the principle of electrical substitution is described. The assembled radiometer possesses high current-voltage sensitivity on the order of 10 V/W, time constant 7 sec, and error in measurement of the emissive power at most 0.3%. It is designed for exact photometric measurements in the visible and infrared bands of the spectrum.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.