Abstract

Cochlear implants are electronic devices used to provide useful hearing sensations to patients with severe to profound hearing loss. Safety and reliability are considered by patients and clinicians when selecting among the different models of cochlear implants. However, comparing reliability estimates from different manufacturers is difficult because of the lack of a standardized approach for classifying and quantifying failure data. This paper develops a classification scheme for cochlear implant failures wherein each failure report is categorized based on the failure site, cause, mode, and mechanism obtained through failure modes, mechanisms, and effects analysis. Failure parameters based on the Weibull distribution are used to quantify the following reliability metrics: B1, B5, B10, and B50 lives. The developed reporting scheme provides clear and quantitative information on failure causes and failure sites for a given implant. This method provides a consistent reporting format for communicating field reliability to regulatory units, clinicians, and patients.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.