Abstract

Cadmium poisoning is a chronic accumulation process, and long-term drinking of even low cadmium content water will cause kidney damage, so an ultra-low detection limit is particularly important. However, at the present stage, the traditional detection method cannot reach a sufficiently low detection limit, the response time is too long, and the cost of detection is very high, so that real-time measurement cannot be realized. Therefore, the traditional cadmium ion detection method has a slow response and an insufficient detection limit. This paper presents a fiber optic cadmium ion sensor functionalized based on an Fe3O4@SiO2@CS magnetic ion imprinting polymer (M-IIP). The sensor is based on the coupling characteristics of the optical microfiber coupler (OMC) cone region to achieve a highly sensitive response to the change in the cadmium ion concentration. M-IIP materials were prepared by surface imprinting polymerization to achieve low cross-sensitivity and thus improve the detection limit of the sensor. The results show that the developed fiber sensor has high specificity and a rapid response to cadmium ions. The experimental limit of detection (LOD) reached 0.051 nM within 0-1 μM with a response time of less than 50 s. Moreover, the proposed fiber cadmium ion sensor exhibits excellent performance in terms of sensitivity, stability, repeatability and biocompatibility.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call