Abstract

Initial feasibility of a spatial frequency domain imaging system was studied consisting of a hand held miniaturized projector and a rigid endoscope. Three wavelengths and two spatial frequencies were used for imaging. The system was calibrated using tissue mimicking phantoms. In vivo imaging was performed on five live mouse tumor models, and the absorption, scattering, hemoglobin oxygen saturation was measured. The initial promising results indicate that the spatial frequency domain imaging can a very useful tool for quantitative wide field tissue evaluation during minimally invasive image guided surgery.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call