Abstract

The aim of the proposed study was to conduct a feasibility study using a flat rectangular (3mm×10mm) MRI compatible transducer operating at 5.3MHz for destroying calcified material in an in vitro model. The proposed method can be used in the future for treating atherosclerosis plaques of the coronary, carotid or peripheral arteries. The system was tested initially on calcium rods. Another test was performed in a hydroxyapatite-polylactide model. A parametric study was performed where the mass of calcified material removed was studied as a function of intensity, pulse repetition frequency (PRF), duty factor (DF) and presence of bubbles. The amount of calcified material removed is directly related to the intensity, PRF and DF. It was found that the presence of bubbles accelerates the removal of calcified material. In order to ensure that pure mechanical mode ultrasound was used, the protocols were designed so that the temperature does not exceed 1°C.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.