Abstract

The electrocardiogram (ECG) represents an essential tool to determine cardiac electrical abnormalities in COVID-19 patients, the effects of anti-SARS-CoV-2 drugs, and potential drug interactions. Smartphone-based heart monitors have increased the spectrum of ECG monitoring however, we are not aware of its reliability in critically ill COVID-19 patients. We aim to evaluate the feasibility and reliability of nurse-performed smartphone electrocardiography for QT interval monitoring in critically ill COVID-19 patients using KardiaMobile-6L compared with the standard 12-lead ECG. An observational comparative study was conducted comparing consecutive KardiaMobile-6L and 12-lead ECG recordings obtained from 20 patients admitted to the intensive care unit with SARS-CoV-2 infection and on invasive mechanical ventilation. The heart rate-corrected QT (QTc) intervals measured by KardiaMobile-6L and 12-lead ECG were compared. In 60 percent of the recordings, QTc intervals measured by KardiaMobile-6L matched those by 12-lead ECG. The QTc intervals measured by KardiaMobile-6 and 12-lead ECG were 428 ± 45 ms and 425 ± 35 ms (p = 0.82), respectively. The former demonstrated good agreement (bias = 2.9 ms; standard deviation of bias = 29.6 ms) with the latter, using the Bland–Altman method of measurement agreement. In all but one recording, KardiaMobile-6L demonstrated QTc prolongation. QTc interval monitoring with KardiaMobile-6L in critically ill COVID-19 patients was feasible and demonstrated reliability comparable to the standard 12-lead ECG.