Abstract

To evaluate the reproducibility of measurement of the fetal left modified myocardial performance index (Mod-MPI) determined using a novel automated system. This was a prospective study of 116 ultrasound examinations from 110 normal singleton pregnancies at 12 + 1 to 37 + 1 weeks' gestation. Two experienced operators each measured the left Mod-MPI twice manually and twice automatically using the Auto Mod-MPI system. Intra- and interoperator reproducibility were assessed using intraclass correlation coefficients (ICCs) and the manual and automated measurements obtained by the more experienced operator were compared using Bland-Altman plots and ICCs. Both operators successfully measured the left Mod-MPI in all cases using the Auto Mod-MPI system. For both operators, intraoperator reproducibility was higher when performing automated measurements (ICC = 0.967 and 0.962 for Operators 1 and 2, respectively) than when performing manual measurements (ICC = 0.857 and 0.856 for Operators 1 and 2, respectively). Interoperator agreement was also better for automated than for manual measurements (ICC = 0.930 vs 0.723, respectively). There was good agreement between the automated and manual values measured by the more experienced operator. The Auto Mod-MPI system is a reliable technique for measuring fetal left Mod-MPI and demonstrates excellent reproducibility.