ABSTRACTAnalytical methods for the determination of inorganic arsenic species have attracted much attention due to the high toxicity of these compounds and related legislative regulations for food. A new method for the separation and quantitation of As(III) and As(V) was developed using ion-exchange membranes and laser-induced breakdown spectroscopy (LIBS). Using the anion-exchange polymer membrane, As(V) was selectively collected on the membrane, and As(III) was filtered through the membrane. The separated As(V) on the membrane was directly determined by LIBS. The As(III) in the filtrate was subsequently oxidized to As(V) and collected by the membrane for LIBS analysis. The detection limit for As(V) was estimated to be 10 mg/kg. The recovery efficiencies for the arsenic species as standards were in the range of 97–99%. This method was applied for the analysis As-spiked water certified reference materials, and the results showed that the recovery for As(V) was 98.9%. This new speciation method is cost-effective, simple, and low labor-intensive for the quantitation of inorganic arsenic.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.