In the present study, we compared Tc-99m HMPAO single photon emission computed tomography (SPECT) brain images obtained using the conventional filtered backprojection (FBP) reconstruction method with the new reconstruction method of conjugate gradients with modified matrix (CGMM). The two methods were employed to generate SPECT images of a brain perfusion study in a 41-year-old patient who had suffered from multiple embolisation of the right hemisphere. Marked improvement in spatial resolution was achieved by CGMM reconstruction as compared to FBP reconstruction; this considerably helped in defining grey matter structures and low-uptake white-matter areas, both in the normal cerebral areas and in the hemisphere that had been affected by stroke. Although the relatively long time still required by the CGMM procedure might be considered as a significant drawback to date, with the increasing diffusion of the new computer facilities the CGMM method can be proposed as a valuable alternative to the standard FBP technique in order to better recognise cerebral areas with different tracer uptake.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.