Abstract

We report the three-dimensional ultrashort-TE (3D UTE) and adiabatic inversion recovery UTE (IR-UTE) sequences employing a radial trajectory with conical view ordering for bi-component T2 * analysis of bound water (T2 *(BW) ) and pore water (T2 *(PW) ) in cortical bone. An interleaved dual-echo 3D UTE acquisition scheme was developed for fast bi-component analysis of bound and pore water in cortical bone. A 3D IR-UTE acquisition scheme employing multiple spokes per IR was developed for bound water imaging. Two-dimensional UTE (2D UTE) and IR-UTE sequences were employed for comparison. The sequences were applied to bovine bone samples (n = 6) and volunteers (n = 6) using a 3-T scanner. Bi-component fitting of 3D UTE images of bovine samples showed a mean T2 *(BW) of 0.26 ± 0.04 ms and T2 *(PW) of 4.16 ± 0.35 ms, with fractions of 21.5 ± 3.6% and 78.5 ± 3.6%, respectively. The 3D IR-UTE signal showed a single-component decay with a mean T2 *(BW) of 0.29 ± 0.05 ms, suggesting selective imaging of bound water. Similar results were achieved with the 2D UTE and IR-UTE sequences. Bi-component fitting of 3D UTE images of the tibial midshafts of healthy volunteers showed a mean T2 *(BW) of 0.32 ± 0.08 ms and T2 *(PW) of 5.78 ± 1.24 ms, with fractions of 34.2 ± 7.4% and 65.8 ± 7.4%, respectively. Single-component fitting of 3D IR-UTE images showed a mean T2 *(BW) of 0.35 ± 0.09 ms. The 3D UTE and 3D IR-UTE techniques allow fast volumetric mapping of bound and pore water in cortical bone. Copyright © 2016 John Wiley & Sons, Ltd.