Abstract

Condensation nucleus counters (CNCs) exhibit slower time response than expected due to mixing effects within the detector.This mixing produces an exponential distribution of delay times with a characteristic mixing time m that ranges from 0.1 s to 0.9 s for commonly used instruments and limits their usefulness for measuring rapidly changing aerosols. Moreover, when used as detectors in the scanning electrical mobility spectrometer (SEMS; also known as scanning mobility particle sizer, SMPS), CNCs limit the speed with which size distribution measurements can be made. In order to overcome this limitation, a new, fast-response mixing CNC (MCNC) has been developed and characterized. The time response of this new detector and TSI Models 3025 and 3010 CNCs has been measured using a spark source to generate an aerosol pulse. The mixing induced response smearing of this new detector, m , of this instrument is 0.058 s, which is significantly shorter than either of the other instruments tested. Its lower detection limit is about 5 nm diameter. The high aerosol flow rate of the MCNC (0.65 l min -1 ), fast time response, and low detection limit make it an ideal detector for SEMS/SMPS measurements. Using this MCNC as a detector for the SEMS, size distribution measurements over the 5 nm to 140 nm range have been made in 3 s with minimal distortion. The size distribution of a coagulation aerosol was effectively recovered by deconvolution with scans as short as 1 s. Uncertainties in the 1 s scans result, in part, from electronics problems in the scanning DMA.