Abstract

The applicability of a monolithic C18-bonded silica column for the rapid HPLC separation of ingredients in medicinal plants and their phytopharmaceutical preparations has been evaluated in the author's laboratory. In this presentation, an existing method for the determination of the iridoid glycoside harpagoside in Harpagophytum procumbens (Devil's Claw) was successfully transferred from a conventional particle-based C18 silica column to a monolithic silica column. The very high porosity of the stationary phase allows chromatography with a much lower backpressure than on conventional columns. Therefore, the flow rate could be easily increased from 0.8 mL/min (particle-based column) to 5 mL/min (monolithic column) and the run-time reduced from 30 to 5 min (that is a reduction about 85%!), without losing any chromatographic resolution of the compound of interest. The amount of harpagoside was measured with the original method on a conventional particle-based silica column and on the adapted method on a monolithic silica column. The statistical mean t-test showed no significant differences of the variances and the means indicating that the fast HPLC method is an acceptable alternative. The shorter analysis time makes the method very valuable for commercial quality control of Harpagophytum extracts and its pharmaceutical preparations.