Abstract
A photometric flow titration based on the redox reaction between KMnO 4 and minoxidil isdescribed. The best titration results were observed at 3.20 x 10 -4 mol L -1 KMnO 4 and 1.00 x 10 -3 mol L -1minoxidil, using the minoxidil solutions as titrant. The flow rate was fixed at 17 mL min -1 and the titrantwas added to the system in aliquots of 500 μL, the color changes were monitored at 550 nm. The methodwas applied to commercial samples and compared with the results from a chromatographic procedure.Recoveries from 97.6 to 102.8 % were observed depending on the sample. Comparison with thechromatographic procedure reveled relative errors of 3.5 - 4.0 %.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.