Abstract

AbstractAmperometry coupled to flow injection analysis (FIA) and to batch injection analysis (BIA) was used for the rapid and precise quantification of ciclopirox olamine in pharmaceutical products. The favourable hydrodynamic conditions provided by both techniques allowed a very high throughput (more than 300 injections per hour) with good linear range (2.0–200 µmol L−1) and low limits of detection (below 1.0 µmol L−1). The results obtained were compared with titration recommended by the American Pharmacopoeia and also using capillary electrophoresis. Good agreement between all results were achieved, demonstrating the good performance of amperometry combined with FIA and BIA.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.