Abstract

A novel method for fast determination of fluoroacetamide, a kind of organic fluorine pesticide, in blood and urine samples was developed with acetamide as an internal standard using gas chromatography/mass spectrometry (GC/MS) after solid-phase microextraction (SPME) technique. The SPME was performed by immersing a PDMS fiber of 100 μm coating thickness in a sample solution for 25 min at 70 °C with (CH 3CH 2) 4NBr to improve the extraction efficiency. After a GC sample injection, the extracted fluoroacetamide was desorbed from the fiber for 4 min to perform the GC/MS detection with a HP-PLOT Q capillary column. The analytical conditions were optimized by examining systematically, the effects of experimental parameters on the ratio of characteristic ion peak areas of fluoroacetamide to acetamide. Under optimal conditions, the ratio was proportional to the concentration of fluoroacetamide ranging from 5.0 to 90 μg/ml with a detection limit of 1.0 μg/ml. The average recovery of fluoroacetamide in blood sample was 92.2%. The established method could be used for the fast and convenient measurement of fluoroacetamide in poisoned sample.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.