Abstract
We study the possibility to use the Cherenkov light for the efficient detection of 511 keV photons with the goal to use it in TOF-PET . We designed and tested two detection modules consisting of PbF2 crystals attached to Planacon MCP-PMT XP85012. Amplified PMT signals are digitized by the SAMPIC module with high readout rate, up to 105 events/s, and a negligible contribution to the time resolution, below 20 ps (FWHM) . We developed a fast 2D scanning system to calibrate the PMT time response and studied in details the timing characteristics of the Planacon PMT. Using a radioactive 22Na source we measured a detection efficiency of 24% for 511 keV photons in a 10 mm thick crystal and a coincidence resolving time of 280 ps. We analyzed the main factors limiting the time resolution of the large-surface detection module and proposed solutions to improve it, which will be tested in our future project.
Submitted Version (
Free)
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have