Abstract
Acoustic droplet ejection (ADE) has become the preferred method for liquid transfer in a variety of applications including synthetic biology, genotyping and drug discovery.Comparing with traditional pipetting techniques, the accuracy and data reproducibility of ADE based liquid transfer are improved, waste and cost are reduced, and cross-contamination is eliminated. The key component in the ADE system is the ultrasound transducer, which is responsible for generating focused ultrasound beam for droplet ejection. However, current ADE systems commonly utilize a single-element focused transducer with a fixed focal length that require mechanical movement to focus on the liquid surface, resulting in reduced liquid transfer efficiency. In this study, we first present a high-frequency annular array transducer for the ADE technology, which enables rapid and dynamic axial focusing to the liquid surface without mechanically moving the transducer, thereby accelerating liquid transfer. Experimental results show that the proposed 10 MHz, 5-element annular array transducer has good dynamic axial focusing ability, and can achieve accurate and stable droplet ejection of nanoliter volume at the designed focal length of 26–32 mm. Our results highlight the potential of the annular array transducer in advancing ADE system for rapid liquid transfer. This technology is expected to be useful in a variety of applications where precise and high-throughput liquid transfer is crucial.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.