Abstract

The effect of blood contamination on the gel-acetylcholinesterase (AChE) test used in the diagnosis of fetal open neural-tube defects was studied with amniotic fluid samples artificially contaminated with fetal or maternal blood in concentrations covering a range exceeding that usually found in clinical practice. Amniotic fluid samples contaminated with maternal blood gave negative gel-AChE results at all concentrations. Contamination with fetal blood yielded positive results if the erythrocyte concentration was greater than about 60 x 10(6) cells/ml. Thus contamination of amniotic fluid with blood is only likely to cause false positive gel-AChE results if this critical concentration is exceeded. Such samples will occur only rarely in clinical practice but when they do the diagnosis should be made with caution.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.