Abstract
Proportional counters filled with tissue equivalent gas mixtures are extremely useful instruments and are being used extensively as sensitive detectors for all types of radiations to measure the energy transferred to small tissue volumes. The linearity of their response is of primary importance. So the investigation and clarification of the physical phenomena taking place in the counter and of the limits within which useful results may be obtained would contribute to a more efficient use and a wider application of these counters. The linearity of response in the dose and in the gas gain has been determined. Linearity in the dose is limited by the total count rate effect, while linearity in the gas gain is limited by secondary processes occurring in the electron avalanche and by the self-induced space charge effect.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.