Abstract

Quantitative differential phase contrast imaging of materials in atomic-resolution scanning transmission electron microscopy using segmented detectors is limited by various factors, including coherent and incoherent aberrations, detector positioning and uniformity, and scan-distortion. By comparing experimental case studies of monolayer and few-layer graphene with image simulations, we explore which parameters require the most precise characterisation for reliable and quantitative interpretation of the reconstructed phases. Coherent and incoherent lens aberrations are found to have the most significant impact. For images over a large field of view, the impact of noise and non-periodic boundary conditions are appreciable, but in this case study have less of an impact than artefacts introduced by beam deflections coupling to beam scanning (imperfect tilt-shift purity).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.