Abstract

The study aimed to determine the factors that impact the size-specific dose estimate (SSDE) for computed tomography (CT) examinations of the chest-abdomen-pelvis and abdomen-pelvis protocols in two clinical radiology practices and evaluate the image quality of these protocols. Imaging parameters, protocols, dose metrics from the CT units and size-related parameters to calculate the SSDE were documented. The image quality of the CT images was assessed using an image subtraction algorithm. The SSDE increased as the volumetric CT dose index (CTDIvol), and the patient's body mass index increased, respectively. Significant differences (p<0.001) occurred between the two hospitals regarding image quality. However, these differences were not indicative of differences in the diagnostic performances for task-based imaging protocols. Different clinical protocols should be reviewed to optimise dose. The inclusion of the pre-monitoring sequence, age of the machine and the scan requisition parameters impacted the SSDEs. Image quality should be assessed to evaluate the consistency of image quality between protocols applied by different CT units when assessing SSDEs.