Abstract

A repeated usable electrode sensor for the measurement of glucose in urine has been developed. For improvement of the selectivity in measuring glucose in urine, we fabricated the Pt electrode coated with multilayered composite of glucose oxidase (GOX) and ascorbate oxidase (AOX). This simple layer structure could deliver more electrons from glucose oxidation without interference from ascorbic acids by first oxidizing the ascorbic acids present in urine. Results showed that our newly developed Pt electrode is capable of quantifying the concentrations of glucose in urine in the range between 0 and 1000 mg/dL, as well as maintaining its detection stability during 200 times of repeated measurements. This repeatable use of the sensor is attributable to the solidified structure of AOX and GOX layer.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.