Abstract

The LAPS device of fast response and high sensitivity, based on electrochemical potential difference, and its system were fabricated for the precise measurement of pH changes and its characteristic were investigated. The electrostatic variation characteristics of LAPS according to the pH changes and parameters in the device were verified through a simulation using LAPS equivalent circuit model. The LAPS device and its system were fabricated on the basis of the result of simulation. The fabricated LAPS system showed linear sensitivity (about 56 mV/pH within the range of pH 2 to pH 11. In order to overcome the defect of general urea sensor (especially slow response time), urease immobilized nitrocellulose membrane was attached on the LAPS and resulted in the very fast response time, 0.29 mV/sec, 0.86 mV/sec at urea concentration of , respectively. And also in order to measure the uranyl ion, the uranyl ion selective sensing membrane with calix[6]arene derivative was used and its sensitivity was 25mV/concentration decade in the wide uranyl ion concentration range of .

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.