Abstract

Image selected in vivo spectroscopy (ISIS) is a volume selection method often used for in vivo (31)P MRS, since it is suitable for measurements of substances with short T(2). However, ISIS can suffer from significant signal contributions caused by T(1) smearing from regions outside the VOI. A computer model was developed to simulate this contamination. The simulation results for the ISIS experiment order implemented in our MR system (ISIS-0) were in agreement with results obtained from phantom measurements. A new extended ISIS experiment order (E-ISIS) was developed, consisting of four "optimal" ISIS experiment orders (ISIS-1 to ISIS-4) performed consecutively with dummy ISIS experiments in between. The simulation results show that contamination due to T(1) smearing is, effectively, eliminated with E-ISIS and is significantly lower than for ISIS-0 and ISIS-1. E-ISIS offers increased accuracy for quantitative and qualitative determination of substances studied using in vivo MRS. Hence, E-ISIS can be valuable for both clinical and research applications.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.