Abstract
Improper automatic exposure control (AEC) termination may result in high overexposures on some radiographic systems. Under AEC, X-ray factors are adjusted automatically to compensate for differences in patient thickness and density. In radiography, AEC is implemented using ionization chambers placed in the film bucky. In this study we deliberately chose incorrect set-up conditions and assessed the response of the AEC system. Two types of incorrect set-up were studied: (1) incorrect selection of bucky radiation detector and (2) simulated misalignment between the X-ray field and light field. The systems tested varied in age from 1 year to in excess of 10 years. In the first test, overexposures of 90 mGy were recorded. Two systems did not meet EC guidelines for improper AEC termination. The second test, misalignment of the X-ray field, was observed to affect the exposure delivered by approximately +/-22%. The maximum dose increase observed, with a chest phantom in the beam, was 165 microGy. Misalignments also resulted in reduced exposures, which may impact on image quality.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.